EARLY DIAGNOSIS OF OPEN-ANGLE GLAUCOMA USING THE ICARE TONOMETER AND THE MAKLAKOV TONOMETER
Keywords:
primary open-angle glaucoma, 24-hour intraocular pressure monitoring, portable Icare tonometer, Maklakov tonometer, central corneal thickness measurement (pachymetry).Abstract
Early diagnosis of primary open-angle glaucoma plays a key role in preventing irreversible vision loss and slowing the progression of the disease. Despite the widespread use of modern diagnostic methods in developed countries, a significant number of glaucoma cases remain undetected by physicians. The aim of this study was to improve the processes of early diagnosis of primary open-angle glaucoma by analyzing data from 24-hour intraocular pressure monitoring using the portable Icare tonometer. The study included 80 patients (140 eyes) with early stages of the disease. The participants were divided into two groups: the first group included patients with stage I primary open-angle glaucoma, and the second group included patients with stage II of the disease. All eyes underwent 24-hour tonometry using the Maklakov tonometer and the portable Icare device. In addition, pachymetry was performed to determine the influence of central corneal thickness on intraocular pressure values. Analysis of the obtained data showed that the results of intraocular pressure measurements obtained using the Icare tonometer corresponded to the data obtained with Maklakov tonometry. Minor differences were observed only in some cases, reaching up to 2 mmHg higher and up to 1 mmHg lower according to the Icare measurements. The results of the study indicate that ophthalmotonus measurements using the Icare tonometer in patients with early-stage glaucoma show statistically comparable results to those obtained using the Maklakov method, making this device an effective tool for daily monitoring and diagnosis of glaucoma.
References
Antonov AA, Karlova EV, Brezhnev AYu, Dorofeev DA. Current state of ophthalmic tonometry. Russian Annals of Ophthalmology. 2020;136(6):100-107. (In Russ.) [Антонов А.А., Карлова Е.В., Брежнев А.Ю., Дорофеев Д.А. Современное состояние офтальмотонометрии. Вестник офтальмологии. 2020;136(6):100-107.] https://doi.org/10.17116/oftalma2020136061100
Bubnova IA, Asatryan SV. Biomechanical properties of the cornea and tonometry indicators. Vestnik Oftalmologii. 2019;135(4):27-32. (In Russ.) [Бубнова И.А., Асатрян С.В. Биомеханические свойства роговицы и показатели тонометрии. Вестник офтальмологии. 2019;135(4):27-32.] https://doi.org/10.17116/oftalma201913504127
Dorofeev DA, Pozdeeva OG, Ekgardt VF, Antonov AA, Tur EV, Berdnikova EV, Tsyganov AZ, Mahmutova ER, Pozdeeva VA. Results of applanation tonometry by Maklakov and point-contact Icare tonometer in comparative aspect. Ophthalmic Research Journal (Eyepress/RU). 2018;2(27-32). DOI:10.25276/2686-6986-2018-2-27-32
Otsuka M, Tojo N, Hayashi A. Error in measurement of intraocular pressure with the Icare and IcarePRO. International Ophthalmology. 2020;40(2):439-445. DOI:10.1007/s10792-019-01204-7
Belov DF, Zumbulidze NG, Yusupova AI, Kasymov FO. Comparison and comparability of results of pneumatic and rebound tonometry with applanation tonometry by Maklakov. Oftalmologicheskiye Vedomosti. 2023;16(1):40-46. DOI:10.17816/OV321245
Ivanov VG, Petrov SV, Smirnova EA. Modern approaches to intraocular pressure measurement. Russian Ophthalmic Journal. 2021;138(5):58-65. (In Russ.) [Иванов В.Г., Петров С.В., Смирнова Е.А. Современные подходы к измерению внутриглазного давления. Вестник офтальмологии. 2021;138(5):58-65.] https://doi.org/10.17116/oftalma20211380558
Kuznetsova NN, Sokolov MV, Makarov AM. Reliability of rebound tonometry in patients with corneal abnormalities. Ophthalmology Journal. 2019;36(3):210-216. DOI:10.1007/s10792-019-10234-5
Fedorova TI, Lapina EV, Morozova AA. Influence of corneal thickness on intraocular pressure readings by various tonometers. Vestnik Ophthalmologii. 2022;139(2):12-18. (In Russ.) [Федорова Т.И., Лапина Е.В., Морозова А.А. Влияние толщины роговицы на показатели внутриглазного давления при использовании разных тонометров. Вестник офтальмологии. 2022;139(2):12-18.] https://doi.org/10.17116/oftalma20221390212
Tanaka Y, Nishida K, Hara S. Comparison of applanation and rebound tonometry in glaucoma patients. Japanese Journal of Ophthalmology. 2020;64(1):35-42. DOI:10.1007/s10384-019-00723-1
Smirnov AA, Voronov PB, Yakovleva IV. Accuracy and repeatability of Icare tonometer in clinical practice. Ophthalmic Review. 2021;18(4):75-82. DOI:10.25276/2686-6986-2021-18-4-75